|
Thomas M. Laue
University of New Hampshire
Department of Biochemistry
Durham, NH
Introduction
Short column sedimentation equilibrium analysis offers many advantages
for researchers wanting quick, sure characterization of the solution properties of
macromolecules. Presented here is a list of when short column methods are an appropriate
choice, a brief theoretical treatment and an overview of short column sedimentation
methods, including ones diagnostic for self-association and nonideality. Short column
methods were first described by Van Holde and Baldwin in 1958 (1),
and centerpieces specifically designed for these methods were described by Yphantis in
1960 (2). Their distinguishing characteristics are (Fig. 1):
1) the parallel rows of four, small, sample viewing (1.2 mm diameter) holes 4 mm apart, 2)
the two rows of larger (2.4 mm) holes used for filling the cell, and 3) the connecting
groove between the fill hole and sample viewing holes.


Figure 1. Eight-channel, short column centerpiece that allows the
viewing of four solution-solvent pairs. Direction of the gravitational field is toward the
keyway. The small holes are observation channels, and the large holes are filling
reservoirs. The holes on the left half hold the solvent only, and those on the right half,
the solvent with macro-molecule. On acceleration the contents of a filling reservoir drain
through the groove into the corresponding observation channel. The holes for the filling
reservoirs are tapered so that they empty completely. For filling, the centerpiece is
positioned in the cell housing as shown in view A, with the top-most holes referred to as
channel A and the bottom-most as Channel D.

Despite their early development, short column sedimentation equilibrium
methods did not enjoy extensive use with the early users of sedimentation equilibrium for
several reasons: 1) manual data acquisition from short columns required nearly the same
amount of time as from longer columns and did not yield as much information per data set;
2) they required higher concentrations of material; 3) they could not be used with the
photoelectric scanner of the Model E; and 4) they often required that a separate
experiment be conducted to measure the sample concentration. However, a great deal has
changed in the intervening years, and all of these limitations have been overcome. Data
acquisition takes seconds now, whereas it could require days in the early '60s. Thus,
there is a distinct advantage to the rapid equilibration available with short columns in
that they provide at least a 20-fold increase in the number of samples that can be
examined in a given period of time. Thus, with a few exceptions, the quantity of
information available from short columns exceeds that from the longer columns. Short
column centerpieces can be used with the photoelectric scanner of the Optima XL-A
Analytical Ultracentrifuge, hence lower concentrations may be used and a separate
concentration determination is not necessary. Finally, improvements in data analysis
(3,4) have made it possible to combine the data from several
short column experiments in order to evaluate the molecular parameters of interest. These
same programs eliminate the need for separate concentration determinations for
interference optics, saving material and time. With these advances, the advantages of
short columns make them the method of first-choice for many routine analyses.
The advantages of short column centerpieces are: 1) only a small volume
(15 microliter vs. 100 microliter) of material is required at moderate
concentrations (0.1 OD), 2) short equilibration times are needed (60-90 min vs.
16-18 h for most molecules), 3) a large number of samples can be examined simultaneously,
and 4) there is minimal radial redistribution of solutes. The first advantage is a clear
benefit when only small quantities of a material are available for analysis. Even more
attractive is the possibility of recovering much of the sample volume (75-90%) after
analysis. When it is desirable to examine a sample over a wide range of buffer conditions,
the first three advantages are of interest. The final advantage is of particular interest
when it is necessary to perform titrations or when characterizing an association between
dissimilar macromolecules (i.e., heteroassociation).
These advantages are so compelling, it is useful to outline when it is
better to use the longer, 3-mm column centerpieces. The longer columns are better to use
when: 1) examining heterogeneous samples, 2) examining low molecular weight materials, 3)
when only very dilute samples are available, and 4) when the sticking of a sample to
centerpiece walls makes it desirable to exploit the lower surface area-to-volume ratio of
the 3-mm column centerpieces.
Below are described some typical applications for short column
centerpieces. These uses have been arranged in sequence ranging from the most simple to
the more complicated. It is important to note, however, that the basic methodology is the
same and that it is the data analysis and interpretation that vary in complexity. This has
the fortunate consequence that even if diagnostic analysis indicates that a chemical
system is exhibiting complicated behavior, it is often only the method of analysis that
needs to be modified, and there is no need to repeat experiments.
Methods
Sample Preparation. There are no special sample preparations
necessary for short column sedimentation equilibrium. Samples used for absorbance
detection should have an optical density between 0.1 and 1.5 at the wavelength of
interest. Samples for Rayleigh interference detection, as used in the Model E, should have
greater than 0.2 fringe (60 microgram/mL) initial concentration. For both optical systems,
it is preferable that the sample be at dialysis equilibrium with the solvent. This can be
achieved either by exhaustive dialysis (1000:1 volume:volume dialysate to sample, with
stirring for 72 h against three changes of buffer) or, more rapidly (10 min) and
conveniently using centrifugal gel filtration (5). Our
experience is that this method is every bit as thorough as exhaustive dialysis, and,
except in cases where sample is lost by binding to the gel matrix or when slow kinetics
prohibit its use (e.g., refolding of a protein from a denaturant), it is the
preferred method.
Cell Handling. Select a centerpiece made from a material that
does not interact with the macromolecules of interest. Short column centerpieces are made
of Kel-F, which is the least likely to absorb H2O or to bind materials from the
solution. However, Kel-F should not be used routinely above 40,000 rpm, since deformation
of the centerpiece is likely.
Because of the high surface-area-to-volume ratio, the centerpiece must
be scrupulously clean to prevent sample contamination. When cleaning, be careful not to
scratch the surfaces of the centerpiece. The surface and holes should be scrubbed with a
nonabrasive detergent (SDS works well) using nonabrasive scrubbers. We have found that
nonwoven cellulose and polyester cleanroom wipes (e.g., DURX 770, Berkshire Corp.,
Great Barrington, MA) are useful for wiping the faces of the centerpieces, while
microswabs (e.g., TX732, Berkshire Corp., Great Barrington, MA) are useful for
cleaning the holes. Thoroughly rinse the centerpiece with distilled, deionized water. For
this latter step, it is useful to equip a small hose with a micropipette tip to direct a
concentrated stream into the holes. Dry the centerpiece with a gentle stream of dry
nitrogen gas, again using a micropipette tip to blow out the holes. It is possible to
include special cleaning steps (e.g., rinsing with EDTA, soaking in protease
inhibitors, use of organic solvents) as long as there is no chemical reaction with the
centerpiece materials and as long as the reagents can be removed without leaving any
residue. Centerpieces should be cleaned and dried immediately after use and should be
stored wrapped in nonshedding paper.
Cell assembly is as described in the manual for the XL-A, except that
the top window is not put into the cell housing until after the centerpiece is filled. For
absorbance detection, the top window should be in a narrow aperture holder (e.g.,
an old Rayleigh-mask window holder), since this holder masks the incident beam, providing
a uniform light pulse over the radial extent of the column. However, adequate results for
most purposes are obtained using the wide-aperture window holders ordinarily used for
absorbance detection. Either window holder may be used with the Rayleigh system.
Serial dilutions are made using the dialysate. In order to eliminate any
dust, samples should be spun in Eppendorf tubes in a microfuge for 5 min. Alternatively,
the samples may be filtered. Failure to do this step usually has no adverse effects unless
large aggregates are present and very low speed analyses (e.g., <3000 rpm) are
performed.
Cell Assembly and Filling. A complete bottom window and a clean
centerpiece are inserted into the cell housing and pressed down using a lint-free tissue
until the window is at the bottom of the housing. A jet of dry nitrogen gas can be used to
blow away any dust on the window prior to inserting the centerpiece and, again, just
before filling. The cell housing is positioned with the keyway towards you so that the
filling grooves form a "V" (Fig. 1A). The larger holes are filled, with the
holes on the left for solvent and those on the right for solution. Solvent-solution pairs
are designated A-D, going from those closest to the center of rotation (A, the furthest
from you during filling) to those closest to the edge of the rotor (D, the closest to you
during filling). Filling is accomplished with a 20-microliter variable pipetter equipped
with a microcapillary tip. The solvent holes are filled with 20 uL of solvent. The
solution holes are first filled with 5 microliters of FC-43* fluorocarbon (used as a base
fluid, below) and 14 microliters of sample, with the most concentrated sample going in
position A and the most dilute in D. Cell assembly is then completed in the normal manner,
except that no filling-hole screws are necessary. Rapidity of filling and assembly is
essential to minimize the effects of evaporation.
Any inert, dense liquid can be used as a base fluid. FC-43 provides a
clear, "square" bottom, and, because its refractive index is close to that of
water, does not produce as much reflection at the interface as do other base fluids. FC-43
itself is extraordinarily inert, though contamination with hydrocarbons can lead to
interactions with solution components. Should this be the case (evidenced by a buildup of
a layer of material at the interface) the FC-43 can be extracted with concentrated H2SO4
followed by exhaustive rinsing with distilled, deionized water.
Short Column Operation. The rotor is loaded and the cells aligned
as described in the XL-A manual. The samples will transfer to the viewing holes at about
5000 rpm. Rotor speeds are chosen as for a longer column equilibrium run such that the
expected value of [[sigma]] is in the range between 2 and 10:
(1)
and is the partial specific volume of the solute, [[rho]] is the
solution density, R = 8.3144 x 107 and T is the temperature in K.
Note that a rough estimate is all that is required, so that for [[sigma]] = 3 an initial
choice of speed can be estimated from:
(2)
where the approximation is valid for aqueous buffer [[rho]] ~ 1,
at temperatures near 25°C and for materials having near 0.7 mL/g. If M(1 -[[rho]])
is completely unknown, it will be necessary to observe the redistribution of the solute at
several rotor speeds, starting at about 2500 rpm. However, to ensure complete transfer of
the sample, the rotor should be brought to 8000 rpm for a couple of minutes before
dropping the rotor speed back to 2500 rpm.
When 14 microliters of solution are used, the column height (from
meniscus to FC-43 interface) should be 700-800 um. A scan should be made of the cell
immediately after reaching speed to ensure sample transfer and to test for leaks. It
should be noted that at very low rotor speeds (<4000 rpm), both the meniscus and base
are distorted by the earth's gravitational field, leading to boundaries with a very broad
appearance. In our experience, solution transfer is not a problem, and leaking, which is
also rare, occurs only at high speeds and between holes along the radial axis.
Estimating the Equilibration Time. One of the greatest advantages
of short columns is the rapid achievement of equilibrium, which, since the time to
equilibrium is proportional to the square of the column height, is about 16 times faster
than for 3-mm columns. The length of time to reach equilibrium can be estimated from the
formulas given by van Holde and Baldwin (1) or can be determined empirically by making
successive overlay scans 10 min apart, looking for any further movement of the
concentration gradient. (A systematic "tail" at the base or meniscus is a good
diagnostic for nonequilibrium.) For most 10-200-kDa proteins, we typically begin testing
for equilibrium after 45 min at speed. Very large molecules will take longer (e.g., E.
coli ribosomes, Mw= 1.46 x 106 require 250-300 min (4)), and viscous solvents will slow down equilibrium by an
amount directly proportional to the viscosity. On the other hand, small molecules and
nonideal solutes tend to reach equilibrium faster than might be expected.
Data Acquisition. Precise, closely spaced data are required for
good analysis. Consequently, the point density should be as high as possible. This means
that for the XL-A absorbance system, a step scan should be used with 0.001-cm step size
and with four repetitions per step. To save time, only those portions of the cell where
light is visible should be scanned. Radial resolution is not a problem with the Rayleigh
optics. The starting and ending radii for channels A through D can be obtained from the
initial scan and, once determined, recorded for use in future experiments. Between 50 and
100 data points should be acquired for each channel.
For absorbance detection, the 4-mm center-to-center spacing of the cell
holes requires that the flash lamp timing be set slightly differently from that ordinarily
used. Optimum results are obtained if a value between 1.65° and1.75° is used as the
spacing between the holes (ordinarily set for 2.5°).The best timing can be determined
using the method described in the XL-A operating manual.
Data Editing. Data from the meniscus and base regions may need to
be eliminated from analysis due to unavoidable refractive effects at the top and bottom of
the cell. If it is desirable to determine average molecular weights with reference to the
loading concentration, it is best if data are removed symmetrically about the midpoint of
the channel. Otherwise, only remove points that are obviously affected by the
discontinuities at the menisci (4).
Methods for Specific Applications
Aside from rapid surveys and cases where there is limited material,
short columns are especially useful for the rapid determination of the association state
of an oligomer, for examining the effects of small ligands on macromolecular associations
and for the study of heteroassociations (i.e., where mixtures of unlike
macromolecules interact). This suitability stems from the fact that component
fractionation is constrained by the short radial distance. There are a few additional
suggestions for using short columns in these applications.
Determination of the Stoichiometry of an Oligomer. This is
determined simply by measuring the molecular weight of the molecule under native
conditions, comparing it to the molecular weight determined under fully denaturing
conditions. The molecular weight for the native material is determined at four different
loading concentrations, one for each of the channels of a short column centerpiece.
Typically, the concentration in channel A is near 1 OD for the absorbance system, or 1
mg/mL for the Rayleigh system, with channels B-D containing a serial
1:2 dilution of the channel A concentration. Only 35 microliters of
solution are needed to make the dilutions: 14 microliters for each channel and 16
microliters for the serial dilution (assuming 2-3 microli are lost to the walls). Cell
loading is described above. When the solutions are run, the redistribution of the solute
will cause the absorbance ranges in the different channels to overlap and to cover an
absorbance range from near zero to 2 OD (depending on rotor speed, of course).
The molecular weight of the denatured material can be determined a
number of different ways. If the oligomer consists of only one type of monomer, a second
cell should be prepared that contains the same sample under fully denaturing conditions.
The two denaturants most commonly used for sedimentation analysis of proteins are 8 M urea
and 6 M guanidine HCl (6,7). Of these, guanidine has been
better characterized for use in sedimentation and, at 6 M, has minimal effect on, even for
glycoproteins. It is important to realize that guanidine can and does
influence(7,8), but that these effects advantageously cancel at 6 M.
Serial dilutions are made with the 6 M guanidine buffer and the second cell loaded as
described above. If the oligomer consists of nonidentical chains, the chains should be
isolated and analyzed separately under denaturing conditions. The cells should be balanced
and loaded into the rotor as described in the manual.
The initial rotor speed should be chosen as described above, using an
estimate(if available) of M for the native material. Should the initial rotorspeed
prove to be too high (e.g., the protein is highly polymerized andthe absorbance
gradient is very steep with material compressed at the bottom),it is worthwhile to stop
the run and shake the cells to redistribute the contents. Simply lowering the rotor speed
will result in an undue wait for re-equilibration. After the data have been acquired at
the lowest rotor speed, the speed should be increased until the ratio of the square of the
rotor speeds is 1.4 or greater (e.g., if the initial rotor speed was 20,000 rpm,
then 24,000 rpm would be a good choice, since [24,000/20,000]2 = 1.44).
Finally, it is useful to examine the solution at a rotor speed high enough that the ratio
of the square of this speed to the first is 3 or greater (e.g., [36,000/20,000]2
= 3.24). The use of multiple speeds will provide data that can be used diagnostically
(below) and will help ensure that suitable gradients will be developed in all channels
under analysis.
Titrations with Small Ligands. Titrations can be made either of
two ways. The first and preferred method is to dialyze a sample against a range of ligand
concentrations. This is done most conveniently using centrifugal gel filtration
(5). Dialysis sets the concentration of free ligand equal to the
concentration of ligand in the dialysate (or in the column). Since sample and reference
have the same free ligand concentration, any optical effects due to the redistribution of
the ligand during sedimentation can be neglected. The second titration method is to simply
add the ligand directly to the sample. With this method only the total concentration of
ligand is known, and systematic errors in the determination of ligand affinities are
possible. In cases where only qualitative results are of interest, this method is
satisfactory, consumes less material and is much faster. Either method works because small
ligands simply do not form significant gradients over the length of a short column. An
estimate of the concentration gradient of the ligand due to sedimentation can be made
using:
(3)
where C0 is the concentration of the ligand and sigma is its
reduced molecular weight (Eq. 1). It can be seen that even fairly large ligands can be
used if C0 and sigm are kept low enough.
Examination of Heteroassociations. Again, the lack of
fractionation of solutes in short columns is an advantage when it is desirable to examine
the interactions between unlike macromolecules. Each macromolecule should be equilibrated
with the solvent (as described above) and analyses of the individual components must be
made so that the behavior of the mixtures can be interpreted properly. It is best to
examine heteroassociating systems at a variety of rotor speeds using several different
loading concentrations and, if possible, covering a range of mole ratios of components
(8,9). The detailed analysis of these sorts of systems is outside
the scope of this review, but specific examples of how these systems can be analyzed
areavailable (8,9).
Overview of Data Analysis
There are three levels of analysis of short column sedimentation data.
The first level is a qualitative assessment of the macromolecular behavior using
diagnostic graphs. At this level the questions are more general, addressing whether or not
the system is homogeneous, whether or not nonideality is significant and assessing whether
or not a mass action association is occurring. At this level of analysis, a reasonable
estimate of the monomer molecular weight is often possible, but quantitative determination
of thermodynamic parameters is not. The second level of analysis consists of combining data
sets from several short column experiments and fitting the ensemble to models using
nonlinear least squares methods (3). Such analyses can provide
estimates of thermodynamic parameters like the monomer molecular weight, association
constants, association stoichiometries and nonideality coefficients. The confidence
interval, i.e., estimates of the precision, for these parameters also can be
determined. However, the accuracy of the analysis depends on using the correct model to
describe the system. If there is only one plausible model, and it fits the data
adequately, then analysis can end at this second level. For many systems, though, a third
level of analysis is necessary. This level consists of testing a variety of plausible
models and choosing which ones adequately describe the data. Often there is no truly
unique model. However, this ambiguity is not due to limitations of short columns nor of
sedimentation, but rather intrinsic to the determination of thermodynamic parameters. A
detailed discussion of nonlinear curve fitting for sedimentation is outside the scope of
this review, but can be found in the literature (3,4), and
some specific applications of nonlinear curve fitting methods to short column data are
listed in Refs. 8-10. The remainder of this note will focus on
the first level of analysis.
An average molecular weight determination is the simplest and most rapid
analysis of short column equilibrium data. The concentration and rotor speed dependence of
these determinations form the basis of the diagnostic graphs described below. Using the
absorbance system, the molecular weight can be estimated from the slope of the graph of
ln(Abs) versus r2/2. These graphs are available from within the
XL-A user interface. Unless heterogeneity or nonideality are particularly severe, graphs
of ln(Abs) versus r2/2 should be sensibly linear. The slope of
this line is sigma (Eq. 1). For the Rayleigh system, which gives only the relative
concentration, sigma should be determined using a nonlinear least squares approach. When
combined with, [[rho]], omega2, R and T, sigma yields the
molecular weight:
(4)
where values for and [[rho]] are readily estimated from sample and
buffer composition (6). If is unknown, the buoyant molecular
weight, Mb, can be used for the diagnostic graphs described below: Mb= M(1
-[[rho]]) = sigm/RTomega2. If the solvent contains a high concentration
of one or more components (e.g., 6 M guanidine or 8 M urea), the effects on of
preferential hydration should be taken into account (6, 7).
In addition to graphical analysis, it is also possible to fit the
concentration distribution data to functions derived from thermodynamic first principles (2,3). For a single ideal thermodynamic species, the absorbance
profile will be an exponential:
(5)
d is the baseline offset and Abs0 is the absorbance at the
reference radial position r0 (2). For a mixture of
ideal species, the resultant curve is the sum of exponentials:
(6)
where Abs0i and sigma i are the reference
concentration and reduced molecular weight for the ith component,
respectively. If a mixture of molecules is present, but the data are treated as though
there were only a single sedimenting species, the sigma determined is an average value for
the components. Under certain conditions, the average value is well-defined and can be
used in the analysis of a chemical system (2-4). The
weight-average molecular weight is defined as:
(7)
where Mi and Ci refer to the molecular weight and
weight concentration of the ith component, respectively. Notice that the
scanner provides absorbances and that the average molecular weight requires weight
concentrations. It is important to realize that if the weight extinction coefficients
differ for the different components, then the molecular weight is not a well-defined
average. Under these circumstances, it is preferable to use the Rayleigh optical system.
The same arguments as above regarding the determination of average
molecular weights hold true, except that if a nonlinear least squares analysis program
is used to fit the data directly and includes d, the baseline offset, as a
parameter, then it is the z-average that is determined as long as all of the
solution column is visible (4,10):
(8)
Notice that the z-average is influenced more strongly by high
molecular weight material than is the weight-average.
Diagnostic Graphs
Two graphs of the molecular weight, one as a function of the total cell
loading concentration and the other as a function of the rotor speed, will serve for the
initial characterization of a system. Based on these graphs, there are five possible
conclusions that can be reached concerning the system: 1) the system is sensibly ideal and
homogeneous, 2) the system is ideal, homogeneous and exhibits a mass-action association,
3) the system is homogeneous and nonideal, 4) the system is homogeneous, nonideal and
exhibits a mass-action association, and 5) the system is heterogeneous with regard to
mass. The latter category does not exclude possibilities of mass-action association or
nonideality. However, while the short column method can be used to detect heterogeneity,
it is not especially good at resolving it. Consequently, aside from describing the
graphical consequences of heterogeneity, the remainder of this discussion will presume
sample homogeneity.
The first diagnostic graph consists of the apparent molecular weight as a function of
the midpoint absorbance of each channel (Fig. 2). This graph will be useful for detecting
nonideality and mass-action associations. Three conditions may be observed: 1) the
molecular weight is constant with changing concentration (absorbance), 2) the molecular
weight decreases with increasing concentration, and 3) the molecular weight increases with
increasing concentration. Should the first result be obtained, it suggests that the
molecule is behaving ideally. A downward curvature to the molecular weight with increasing
concentration indicates thermodynamic nonideality. Useful information may be available
from this concentration dependence, and methods for analyzing nonideality are available
(11). Increasing molecular weight with increasing concentration is indicative of a
mass-action association and will demand a more detailed analysis. Often at high
concentrations the reaction is driven to an end point (i.e., the association has an upper
limit), in which case the stoichiometry of the largest oligomer may be determined. In
other cases the association may be indefinite or experimental complications (e.g.,
nonideality or limited solubility or availability) may make it difficult to determine an
upper limit for the oligomer size. Even in the case of a limited association, the
stoichiometry (N) becomes difficult to determine as 1/N approaches the precision of the
molecular weight estimate (typically 1-3%).


Figure 2. Diagnostic graph providing a qualitative characterization of the solution
behavior of macromolecules. All data were acquired at a constant temperature between 20 and
25º C. This figure shows the apparent molecular weight as a function of cell loading concentration.
Because of the disparity in the monomer molecular weights for the materials shown here,
the y-axis is presented as M/M1 where M1 is the "monomer"
molecular weight. () shows behavior typical for an ideal, nonassociating
material. Presented here are the data for bovine articular cartilage link protein
(M1 = 260,000) in 0.15 M NaCl, 50 mM Tris (pH 7), 5 mM EDTA. This same
sample in 6 M guanidinium buffer gives a monomer molecular weight of 43,000,
demonstrating that link protein is a hexamer under nondissociating conditions.
(■) shows the behavior typical of a nonideal macromolecule. Presented here are the
data for the glycosylaminoglycan chains of bovine articular dermatan sulfate
proteoglycan I (M1 = 26,000) in the same buffer. These chains are highly
charged, leading to the nonideality observed here. The identical monomer molecular
weight is obtained when the chains are sedimented in 6 M guanidine-containing buffer
indicating that they are monomeric in associative buffer. (▲) presents the
behavior typical of a mass-action association. The sample in this case is MyoD-bHLH
(M1 = 16,000) in 100 mM NaCl, 20 mM acetate, 50 mM Tris, pH 7.6, 2 mM
ß-mercaptoethanol. These data suggest that MyoD-bHLH undergoes an apparent monomer-dimer
equilibrium. However, mass spectroscopy, cDNA sequence analysis and sedimentation in 6 M
guanidium buffer reveal that the true monomer molecular weight is 8000, so the reaction
observed here is actually a dimer-tetramer equilibrium.

Graphs of the apparent molecular weight as a function of rotor speed are useful for
detecting heterogeneity (Fig. 3). They also are important for determining the reversibility
of a mass-action association (3). For a homogeneous, noninteracting
sample under all conditions, the molecular weight should be independent of rotor speed
(3, 4). The same is true for a homogeneous mass-action association, as
long as the concentration gradient can be monitored all the way from the meniscus to the
base. On the other hand, a systematic decrease in the molecular weight as the rotor speed
is increased is diagnostic for sample heterogeneity. Such heterogeneity often is
accompanied by a band of material gathering at the FC-43 layer at higher rotor speeds.
Because the hydrostatic pressure developed in a short column is quite small, even at high
rotor speeds (a few atmospheres), pressure-dependent dissociation is an unlikely cause of
decreasing molecular weight with increased rotor speed, thus simplifying the
interpretation. Should heterogeneity be suspected, further fractionation of the sample
(e.g., by size-exclusion chromatography) is required. For a heterogeneous mass-action
association, the apparent molecular weight can exhibit rotor-speed dependence depending on
how tight the oligomerization is and whether all reactants are present in the
stoichiometrically correct amounts. Under these circumstances, these graphs are not useful
and more detailed analysis will be required (see above).
A few cautions are in order. First, macromolecular association and thermodynamic
nonideality produce opposite effects on the concentration dependence of the molecular
weight. Cases have been reported in which what appears to be ideal behavior actually
results from compensating effects of association and nonideality. If this is of concern,
it can be tested by reexamining the solution under slightly different conditions of pH or
salt (11). Second, these graphs are better thought of as qualitative
guides and should not be used to estimate association nonideality or molecular weights.
For example, mass-action dimer formation will yield an Mz that is 80% of dimer
at the concentration equaling the dissociation constant (Kd). Therefore, using
the concentration at which Mz is one-half that of the dimer would yield a
serious underestimate of Kd. These parameters are best determined using the
curve fitting methods described above (3, 4). On the other hand, curves
like those in Fig. 2 certainly can lead to plausible models for such curve fitting.


Figure 3. Diagnostic graph providing a qualitative characterization of the solution
behavior of macromolecules. All data were acquired at a constant temperature between 20
and 25º C. This figure shows the apparent molecular weight (presented as M/M0,
where M0 equals the molecular weight determination at the lowest rotor speed)
as a function of rotor speed for a homogeneous sample (■) and a heterogeneous sample
(). The x-axis is the ratio of the RPM to the lowest RPM0 rotor speed used.
The homogeneous sample is LexA (0.8 mg/mL) in 50 mM KCl, 20 mM Tris, pH 7.4, 10% sucrose,
2.5 mM MgCI2, 1.5 mM CaCl2, 0.5 mM EDTA and 1 mM dithiothreitol, and analyzed at rotor
speeds of 22,000, 28,000 and 34,000 RPM. LexA (M1 = 23,000) undergoes a
reversible monomer-dimer equilibrium such that at this concentration M0 35,000.
The lack of any apparent change in the molecular weight with increasing rotor speed is
thermodynamic proof that the equilibrium is fully reversible (3, 12). The
heterogeneous sample is core protein from bovine articular cartilage dermatan sulfate
proteoglycan I (1.1 mg/mL) in 0.1 M NaCl, 50 mM Tris, pH 7.6, and 5 mM EDTA analyzed at
16,000, 20,000, 24,000 and 28,000 rpm. The pronounced decrease in apparent molecular weight
with increased rotor speed in this case is indicative of sample heterogeneity.

Sample Recovery
One of the attractive features of sedimentation analysis is that it is nondestructive
to the sample, and it is possible to recover much of the sample from a short column
centerpiece. To do this, the centerpiece is held in place using a fill-hole screw while the
top window is removed. To facilitate removing the window, we have drilled two small holes
into the top rim of the window holder so that screws may be inserted to serve as handles.
(Similar modifications are available from Beckman on a special request basis.) There is
inevitable loss of liquid to the window as it is lifted off, but the majority of it
remains in the holes. Once the window is off, the samples may be removed using a
micropipette equipped with a microcapillary tip. Typically 70% or so of the initial 14 µL
is recovered. Procedures for optimizing sample recovery have not been described, so it is
possible that even greater quantities could be retrieved.
Summary
Short column sedimentation equilibrium provides a rapid and sure method for
characterizing the solution behavior of macromolecules. The small quantity of material
required and the ability to recover the sample unharmed make it an attractive method for
molecular biology. Short columns are particularly useful for conducting rapid surveys of
the environmental influences on macromolecular association.
Acknowledgment
The work presented here was supported by a National Science Foundation grants
DIR 90-02027. The author thanks Daryl Lyons and Jun Liu for providing the experimental
data presented here; Dr. Lawrence Rosenberg for providing articular cartilage link protein,
dermatan sulfate proteoglycan core protein and GAG chains; Dr. John Little for providing
LexA; and Dr. Rachel Klevit for providing MyoD-bHLH peptide.
References
- van Holde, K. E., Baldwin, R. L. Rapid attainment of sedimentation equilibrium.
J. Phys. Chem. 62, 734-749 (1958).
- Yphantis, D. A. Rapid determination of molecular weights of peptides and proteins.
Ann. N. Y. Acad. Sci. 88, 586-601 (1960).
- Johnson, M. L., Correia, J. J., Yphantis, D. A., Halvorson, H. R. Analysis of data from the analytical ultracentrifuge by nonlinear least-squares techniques.
Biophys. J. 36, 575-588 (1981).
- Correia, J. J., Yphantis, D. A. Equilibrium sedimentation in short columns.
Analytical Ultracentrifugation in Biochemistry and Polymer Science. Edited by S. Harding and A. Rowe. London, Royal Society for Chemistry, (in press).
- Christopherson, R. I., Jones, M. E., Finch, L. R. A simple centrifuge column for desalting protein solutions.
Anal. Biochem. 100, 184-187 (1979).
- Laue, T. M., Shah, B. D., Ridgeway, T. M., Pelletier, S. L. Computer-aided interpretation of analytical sedimentation data for proteins.
Analytical Ultracentrifugation in Biochemistry and Polymer Science. Edited by S. Harding and A. Rowe. London, Royal Society for Chemistry, (in press).
- Durschlag, H. Thermodynamic Data for Biochemistry and
Biotechnology, Chapter 3, p. 45. Edited by H.-J. Hinz. New York, Springer-Verlag, 1986.
- Luckow, E. A., Lyons, D. A., Ridgeway, T. M., Esmon, C. T., Laue, T. M. Interaction of clotting factor V heavy chain with prothrombin and prethrombin 1 and role of activated protein C in regulating this interaction: analysis by analytical ultracentrifugation.
Biochemistry 28, 2348-2354 (1989).
- Olsen, P. H., Esmon, N. L., Esmon, C. T., Laue, T. M.
Ca2+ dependence of the interactions between protein C, thrombin, and the elastase fragment of thrombomodulin . Analysis by ultracentrifugation.
Biochemistry 31, 746-754 (1992).
- Arakawa, T., Yphantis, D. A. Molecular weight of recombinant human tumor necrosis
factor-a. J. Biol. Chem. 262, 7484-7485 (1987).
- Wills, P. R. Thermodynamic non-ideality and sedimentation analysis.
Analytical Ultracentrifugation in Biochemistry and Polymer Science. Edited by S. Harding and A. Rowe. London, Royal Society for Chemistry, (in press).
- Squire, P. G., Li, C. H. Adrenocorticotropin (ACTH).
XXIII. A sedimentation study of the state of aggregation of ovine pituitary ACTH in acidic and basic solutions.
J. Am. Chem. Soc. 83, 3521-3528 (1961).
|