Janet Condino
Beckman Coulter, Inc.
Sedimentation equilibrium provides a powerful method for determination
of the quaternary structure of associating macromolecules, as well as for study of the
stoichiometry of heterogeneous binding reactions. Both types of study rely upon a
determination of the average molecular weights of species allowed to equilibrate in the
sample cell of the analytical ultracentrifuge rotor. In order to better understand the
analyses employed to study associating systems, it is useful to examine the methods used
for simple determinations of molecular weights. Although several methods exist for
obtaining molecular weights that appear to be simpler and less time consuming, the
versatility and power of sedimentation analysis endows the analytical ultracentrifuge with
a unique position in the array of alternative techniques.
With respect to molecular weight determinations, the advantages of
sedimentation analysis are as follows. The wide range of rotor speeds makes possible the
study of molecules as small as sucrose and as large as a virus of 50 million daltons. For
polydisperse samples, average molecular weights can be obtained as well as some measure of
the molecular weight distribution. The usual methods employed do not require any
assumptions about shape or hydration of the solute under study. Furthermore, the
measurements made in the ultracentrifuge are primary. Unlike commonly used methods that
require the use of standards, no calibration with materials of known molecular weight are
required. This is particularly important when the molecule being analyzed and the standard
molecules behave somewhat differently due to subtle differences in physical parameters.
Two general methods for determining molecular weights using analytical
ultracentrifugation have been used: the sedimentation and diffusion method and the
sedimentation equilibrium method. The first of these, widely used in the past, has been
largely replaced by the sedimentation equilibrium method. It is of historic interest,
however, and will be covered in a brief section at the end.
The most common method of molecular weight determination is
sedimentation equilibrium. This technique depends on the measurement of the solute
concentration distribution in the sample cell when equilibrium conditions have been
attained, that is, when the distribution of solute due to sedimentation is balanced by the
distribution due to diffusion.
While the theory of sedimentation equilibrium can be viewed from a
kinetic standpoint(1) or from a thermodynamic one,(2) the equations relating the final concentration
distribution in the cell to the molecular weight of the solute are the same in both cases.
In the kinetic treatment, the equilibrium condition is seen to be achieved when the rate
of transport of material in one direction due to sedimentation is exactly balanced by the
rate of transport in the other direction due to diffusion. In the thermodynamic treatment,
equilibrium is reached when the total potential (the sum of the chemical potential and the
centrifugal potential) is independent of time and of position in the cell. Strictly
speaking, the final equilibrium condition is never actually attained because it would
require infinite time. In practice, it is possible to achieve in a reasonable time
interval a condition that is, for computational purposes, equivalent to the theoretical
equilibrium condition. This practical equilibrium condition is one in which the
concentration distribution in the centrifuge cell does not change with time within the
experimental error in measuring the distribution.
Compared to the sedimentation and diffusion method mentioned above, the
equilibrium technique has several advantages: (a) it has a secure theoretical basis in
thermodynamics so that fewer assumptions are required in order to apply the method to
experimental data, (b) it is simpler experimentally, and (c) it is more accurate.
Sedimentation equilibrium is especially preferred over sedimentation and diffusion in
studies of polydisperse solutes, or for solutions containing high concentrations of salts
or denaturing agents.
Conventional Sedimentation Equilibrium Method
"Conventional" is used here to distinguish this method from
approach-to-equilibrium, short column, and meniscus depletion methods. Conventional
sedimentation equilibrium experiments are defined as those experiments that utilize a
fluid column length of 3 to 5 mm in the centrifuge cell, with rotor speeds sufficiently
low so that solute concentration at the meniscus is not zero, and with the entire contents
of the cell at sedimentation equilibrium. In order to compute molecular weights, the
concentration of the solute throughout the cell must be determined when equilibrium is
reached. It is therefore important to know at least an approximate time to reach
equilibrium (the transient time), and an appropriate rotor speed. If an approximate
molecular weight for the solute is known, then operating speeds and transient times can be
estimated from equations available in the literature.(2, 3)
However, the final judgment as to when equilibrium has been reached is based on
experimental factors: once presumed equilibrium has been reached, take scans several hours
apart at intervals of approximately one-fifth the estimated equilibrium time and compare
the data. When the difference from scan to scan is the same as the noise in the data,
assume equilibrium has been reached.
The criterion usually given for the optimum rotor speed for 3-mm
solution columns is one that will result in a solute concentration at the bottom of the
cell that is four times the concentration at the meniscus. This rule is applicable only
for homogeneous, ideal solutes and should be considered to be a minimum speed. Figure 1
can be used to obtain an estimate of an appropriate operating speed if an approximate
molecular weight is known. If no information about the sample is available, make a
separate sedimentation velocity run(4) and determine
an approximate sedimentation coefficient; then use Figure 1. This first estimate of
operating speed can be off as much as 50% without any serious consequences.

Figure 1. Nomogram for selecting an operating speed for
a sedimentation equilibrium run.

Transient time can be reduced by a technique known as overspeeding.(5, 6) This is a simple procedure in which a rotor is
initially run at a speed higher than the equilibrium speed to promote rapid redistribution
of the solute. The rotor speed is then decreased to the selected speed and the solute is
allowed to reach its final equilibrium distribution. During the initial period, bulk
solute is transported by sedimentation toward the bottom of the cell, so that the quantity
of material in the upper and lower halves of the cell grossly approximates that required
for the final equilibrium condition. The rotor is then decelerated to a speed lower than
the equilibrium speed to allow diffusion of the steep gradients formed during the initial
period. Finally the rotor is accelerated to the equilibrium speed and allowed to run until
the required solute distribution is attained. A detailed procedure for computing the
required speeds and times of operation for each step has been described by Hexford,
Radford, and Beams.(5) In practice, two speeds are
adequate: an initial overspeeding period followed by reduction to the equilibrium speed.
Although this is a valuable technique, it is not without a hazard: if the concentration of
the solute at the bottom of the cell becomes so high that it forms a gel-like precipitate,
it may not redissolve when the speed is reduced. The following method is the best general
procedure for most samples. The fluid column is sufficiently short so that the time
required for attainment of the equilibrium condition is not excessive (most proteins will
require 24 hours or less). At the same time, the column length is sufficiently long to
allow accurate determination of molecular weights and assessments of possible
heterogeneity or nonideality. If the sample is polydisperse, then the point weight-average
molecular weights across the cell, as well as the weight-average and z-average
molecular weights for the whole sample can be calculated from the same experimental data. (note 1) For reacting systems, the same procedure can be
used and extended to approximate the equilibrium constants for the system. For studies of
associating systems, longer column heights are preferred because the greater amount of
data obtained will result in higher precision.
Recommended Procedure
The best general references are Richards and Schachman,(7) LaBar and Baldwin,(8)
and Van Holde.(3)
Sample and cell: use a 12-mm double-sector cell, quartz window, and a
double-sector centerpiece. Torque to 120 inch-pounds. With the screw-ring end facing you
and with the filling holes up, place 110 µL of sample diluted in buffer or salt solution(9) in the right-hand sector of the cell. Absorbance of the
sample solution should be close to 0.2 to 0.4 at the selected wavelength. In the left-hand
sector, place 120 µL of solvent, preferably the dialyzate of the sample solution. Adjust
the weight of the appropriate counterbalance to within 0.5 g of the weight of the filled
cell, but not more than the weight of the cell. Align the cell and counterbalance
carefully.
Optionally, FC-43(note 2) oil
can be placed in the bottom of double-sector centerpieces. In addition to lifting the
aqueous sample solution off the cell bottom (necessary for some centerpieces), FC-43 has
the advantage of a refractive index very close to that of water. Consequently, FC-43 won't
reflect light at the cell bottom to the same extent that the centerpiece material will,
and data collected in this region may be more reproducible. Place 30 µL of FC-43 in the
right-hand sector of the cell followed by 80 µL of sample in dilute buffer or salt
solution. (The FC-43 may contain impurities that cause the sample to aggregate, in which
case it can be omitted.)
If you choose to overspeed to reduce the time to equilibrium, initially
use a speed 1.5 times the final equilibrium speed (but not in excess of the speed limit of
the rotor) and run at that speed for about one-fifth of the total estimated equilibrium
time. Decrease to the final equilibrium speed and run at that speed for the remainder of
the estimated transient time. When equilibrium has been reached (see above), stop the run
and analyze the data.

1 A weight-average molecular weight is
defined as
and a z-average molecular weight is defined as
, where Ci is the concentration of component i,
and Mi is the molecular weight of component i.
2 FC-43 oil (3M Company) is a dense (1.92
g/mL), immiscible fluorocarbon.(10)

Data Analysis
For any macrosolute with negligible nonideality, the concentration
distribution in the centrifuge cell at equilibrium is described by
where
| Cr |
= |
concentration of macrosolute at any radial distance |
| Cr0 |
= |
concentration of macrosolute at the
reference radial distance r0 |
| omega |
= |
rotational velocity in radians/s |
| v-bar |
= |
partial specific volume of macrosolute |
| rho |
= |
density of solution |
| R |
= |
gas constant |
| T |
= |
absolute temperature |
| M |
= |
molecular weight |
(In practice, this equation is often written with absorbance, A,
substituted for concentration, C, assuming by Beer's law that absorbance is
proportional to concentration.) Rearranging the equation yields a way to calculate
molecular weight:
Plotting ln(Cr) versus r2 will give
a line with a slope that equals M(1 - v-bar rho), the buoyant molecular weight of
the macrosolute.
Complications arise, however, when this plot is not a straight line. A
downward curving plot, indicating that the molecular weight decreases with increasing
concentration, is an indication that the solution is nonideal, and that the molecular
weight exhibits concentration dependence. An upward curving plot, on the other hand, is an
indication that the macrosolute is polydisperse. It is a mixture of molecular weights
either because it is impure or because sample material has aggregated. In this case dln(Cr)/dr2
yields an average molecular weight that, providing all macrosolute species have the same
partial specific volume, gives the weight-average molecular weight, Mw.
If the polydispersity is a reflection only of a self-associating system
where aggregates are in reversible equilibrium with monomers, a weight-average molecular
weight at any point in the cell will depend only on the concentration of the macrosolute
at that point. Plots of Mw versus solute concentration, when runs are
done at different rotor speeds, should all overlap within experimental errors. If, on the
other hand, the polydispersity is not due to a self-association, plots of Mw
and Cr at different rotor speeds will not overlap.
Using the linearity of the ln(Cr) and r2
plot as an assessment of homogeneity is not without pitfalls. One is the inability of the
eye to detect minor deviations. For example, the curvature caused by dimerization of as
much as 10% of the solute will generally be undetectable. A more subtle effect exists when
a sample exhibits both concentration dependence and polydispersity. One may cancel out the
other, and the plot of ln(Cr) and r2;would be a
straight line, although the sample is neither homogeneous nor ideal. Experiments at other
concentrations and/or other rotor speeds will provide a test for both effects. At higher
speeds the effects of heterogeneity should become more obvious. At lower concentrations
the effects of nonideality will decrease in an almost linear fashion. If the sample is
known to be homogeneous by some other means, then a curving plot will be due only to
nonideality. In general, if the plot is a straight line and the speed sufficient to assure
a large change in concentration from the top to the bottom of the cell (Cr/Cr0
of 4 to 10, for example), then one can be reasonably sure that the molecular weight value
obtained is independent of concentration.
While the plot of log (Cr) versus r2
is a well-established approach to the analysis of sedimentation equilibrium data, there
are limitations, as illustrated above, to this method of analysis and one may wish to use
a more sophisticated approach to data analysis.
Molecular Weights from Sedimentation and Diffusion Data
The sedimentation and diffusion method (s/D) is based on the
relationship between the sedimentation coefficient and the molecular weight of the
molecule. The well-known Svedberg equation(2, 10) is
used for this calculation.
The inherent limitation of this method of determining molecular weights
is the need to know not only the partial specific volume of the solute and the density of
the solution, but also the sedimentation and diffusion coefficients of the solute. These
values, determined at a single finite concentration, can be used for the computation of an
apparent molecular weight value providing that the experimental conditions of solvent,
concentration, and temperature are identical for both determinations. If s and D
values are used, then theoretically the value of the density used in the equation should
be the density of the sample solution in water at the experimental concentration, at
20°C. In practice, the value for the density of water gives results of sufficient
accuracy. The value for T used in this case is 293.2°K.
The best results are obtained using values of s and D that
have been determined at a series of concentrations in the same solvent at the same
temperature and extrapolated to zero concentration. These values of s and D
(not necessarily corrected to water at 20°C) are then used in the Svedberg equation,
along with the density of the solvent and the experimental temperature. The value of the
molecular weight obtained by this method is that of the anhydrous molecule, providing the
v-bar used is the anhydrous value.
Compared to the sedimentation equilibrium technique, the s/D
method has an advantage when paucidisperse solutes are studied because molecular weights
can be determined for each component separable by sedimentation velocity (providing the
essential diffusion data are available). The same advantage applies in special cases of
mixtures of molecules of the same size and different shape or of the same shape and
different size. In the study of some molecules that are too large for the sedimentation
equilibrium technique, the s/D technique may be used advantageously. In particular, large
nucleic acids cannot be studied by sedimentation equilibrium methods, so molecular weight
estimations for these materials are generally made based on sedimentation coefficients.
When a macromolecule, which has been shown to be pure by other methods,
demonstrates heterogeneity with respect to mass by sedimentation equilibrium, further
experiments are needed to distinguish a self-associating system from one where the
polydispersity is due to contaminants. For associating systems, the monomer molecular
weight as well as the degree of aggregation can be estimated. Consult the literature as
suggested below for further uses of the sedimentation equilibrium technique:
- Computation of point average molecular weights(11)
- Estimation of dissociation constants for reacting systems(12, 13)
- Study of molecular weight distributions, including synthetic polymers in
organic systems(14-20)
In addition to the conventional approach described here, variations such
as approach-to-equilibrium(4, 21-24) and the short
column technique(14, 25-27) provide, for appropriate
samples, the power of the traditional method with the advantage of reduced run times.
References
- Svedberg, T., Pederson, K. O. The Ultracentrifuge, p. 5. Oxford
and Clarendon Press, 1940. (Johnson Reprint Corporation, New York)
- Ibid, p. 16.
- Van Holde, K. E. Sedimentation equilibrium. Fractions, Palo Alto,
Calif., Beckman Coulter, Inc., 1967, No. 1, p. 1.
- Furst, A. Overview of sedimentation velocity for the Optima
XL-A Analytical Ultracentrifuge. Applications Data DS-819.
Palo Alto, Calif., Beckman Coulter, Inc., 1991.
- Hexner, P. E., Radford, L. E., Beams, J. W. Achievement of sedimentation
equilibrium. Proc. Natl. Acad. Sci. 47, 1848 (1961)
- Richards, E. G., Teller, D. C., Schachman, H. K. Ultracentrifuge studies
with Rayleigh interference optics. II. Low-speed sedimentation equilibrium of homogeneous
systems. Biochemistry 7, 1054 (1968)
- Richards, E. G., Schachman, H. K. Ultracentrifuge studies with Rayleigh
interference optics. I. General applications. J. Phys. Chem. 63, 1578 (1959)
- LaBar, F. E., Baldwin, R. L. A study by interference optics of
sedimentation in short columns. J. Phys. Chem. 66, 152 (1962)
- Condino, J. Sample preparation for analytical ultracentrifugation in the
Optima XL-A. Technical Information LXL/A-AN-001. Palo Alto, Calif.,
Beckman Coulter, Inc., 1992.
- Schachman, H. K. Ultracentrifugation, diffusion, and viscometry. Methods
in Enzymology, Vol. 4, p. 32. Edited by S. P. Colowick and N. O. Kaplan. New York,
Academic Press, 1957.
- Yphantis, D. A. Equilibrium ultracentrifugation of dilute solutions. Biochemistry
3, 297 (1964)
- Adams, E. T., Fujita, H. Sedimentation equilibrium in reacting systems.
Ultracentrifugal Analysis in Theory and Experiment, p. 119. Edited by J. W.
Williams. New York, Academic Press, 1963.
- Adams, E. T., Williams, J. W. Sedimentation equilibrium in reacting
systems. II. Extension of the theory to several types of association phenomena. J. Am.
Chem. Soc. 86, 3454 (1964)
- Pedersen, K. O. Details of methods of measurement. The
Ultracentrifuge, p. 237. Edited by T. Svedberg and K. O. Pederson. Oxford, Clarendon
Press, 1940. (Johnson Reprint Corporation, New York)
- Wales, M., Rehfeld, S. J. Molecular weight distribution by velocity
ultracentrifugation. J. Polymer Sci. 62, 179 (1962)
- Wales, M., Williams, J. W., Thompson, J. O., Ewart, R. H. Sedimentation
equilibrium of polydisperse non-ideal solutes. Experiment. J. Phys. Colloid Chem. 52,
983 (1948)
- Wales, M., Adler, F. T., Van Holde, K. E. Sedimentation equilibria of
polydisperse non-ideal solutes. VI. Number-average molecular weight and molecular weight
distribution functions. J. Phys. Colloid Chem. 55, 145 (1951)
- Donnelly, T. H. The study of limited molecular weight distribution by
the use of equilibrium ultracentrifugation. J. Phys. Colloid Chem. 64, 1830 (1960)
- Billmeyer, F. W., Jr. Characterization of molecular weight
distributions in high polymers. J. Polymer Sci., Part C, 8, 161 (1965)
- Scholte, T. G. Molecular weights and molecular weight distribution of
polymers by equilibrium ultracentrifugation. I. Average molecular weights. J. Polymer
Sci., Part A-2, 6, 91 (1968)
- Scholte, T. G. Molecular weights and molecular weight distribution of
polymers by equilibrium ultracentrifugation. II. Molecular weight distribution. J.
Polymer Sci., Part A-2, 6, 111 (1968)
- Archibald, W. J. A demonstration of some new methods of determining
molecular weights from the data of the ultracentrifuge. J. Phys. Colloid Chem. 51,
1204 (1947)
- Tanford, C. Physical Chemistry of Macromolecules, p. 384. New
York, John Wiley & Sons, 1963.
- Trautman, R. Ultracentrifugation. Instrumental Methods of
Experimental Biology, p. 255. Edited by D. W. Newman. New York, The Macmillan Company,
1964.
- Van Holde, K. E., Baldwin, R. L. Rapid attainment of sedimentation
equilibrium. J. Phys. Chem. 62, 734 (1958)
- Yphantis, D. A. Rapid determination of molecular weights of peptides
and proteins. Ann. N.Y. Acad. Sci. 88, 586 (1960)
- Laue, T. M. Short column sedimentation equilibrium analysis. Technical Information DS-835. Palo Alto, Calif.,
Beckman Coulter, Inc., 1992.
|