 |
High-Throughput Genomic DNA Isolation from Cell Cultures and Mouse Tails
on the Biomek® FX Liquid Handling System using Promega's Wizard* SV 96
Genomic DNA Purification System
Hobert Wai, Ph.D.**, Matthew Cu**, Dana P. Campbell**, and Keith Roby**
Beckman Coulter, Inc.
**Indianapolis, IN and ***Fullerton, CA
Terri Grunst and Daniel Kephart
Promega Corporation –
www.promega.com
For additional information try our Maintenance
and Troubleshooting section or post a question to our Biomek
User Forum. New members can subscribe to our User Forum here.
Click here to download a printable
PDF version of the following:
| Problem |
Possible Cause |
Solution |
| Deck mapping and setup
difficulties. |
- Calling up the wrong deck.
Not enough ALP positions.
|
- Refer to Beckman Coulter Application Information Bulletin
T-1953, “Deck Mapping on the Biomek® FX.”
|
|
Low DNA yield
(i.e., low A 260). |
- Sample contained too few cells.
|
- Obtain fresh new cell samples.
- Check to see if cell count is within the recommended
range.
|
- Tissue lysate is stored at -20 or -70°C.
|
- For optimal yield, purify the DNA as soon as the lysate
is prepared. Generally, if the lysate has been frozen it
may have a decreased amount of genomic DNA.
|
- Sample subjected to many freeze-thaw cycles.
|
- Freezing and thawing samples repeatedly could result in
DNA degradation. Use fresh samples whenever possible.
|
- Tissue culture cells were low in genomic DNA.
|
- Genomic DNA yield varies
depending on the number of cells used for the isolation.
Increase the starting amount processed to a maximum of 20
mg of tissue or 5 X 106 tissue culture cells.
|
- Sample did not receive any or enough Wizard SV Lysis
Buffer.
|
- Process lysates at 55°C
incubation. If samples have
cooled, it may be difficult to purify due to viscosity.
Incubate at 55°C for 1 hour and continue with the
purification.
|
- Lysed cells or lysates were not re-suspended completely.
|
- Increase number of sample mixes to ensure re-suspension
before transferring to the DNA filter plate.
|
- No ethanol or incorrect volume of ethanol added to the
Wizard SV Wash Solution.
|
- Prepare the solutions as instructed in Promega’s
Technical Bulletin No. 303 Sections III.A and IV.A
before beginning the procedure.
|
|
|
- Check vacuum pump for correct pressure (15 inches of
Hg).
- Check vacuum lines and seals for leakage.
|
|
| Inaccurate elution
volume. |
- There is not enough nuclease-free water.
|
- Make sure to have enough elution water.
- Allow 10 mL overage when adding to the reservoir.
|
- The aspirating height is not adjusted to the volume of
the nuclease-free water.
|
- Adjust the aspirating height
according to the amount of
nuclease-free water.
|
- Vacuum pump is not working properly.
|
- Check to see that vacuum pump is set to the required pressure.
A vacuum pressure of >15 inches of Hg is recommended.
- Make sure that vacuum hoses are in good conditions (i.e.,
no holes or deterioration) and correctly connected—see
the Biomek® FX User’s Manual #719452.
- Check the manifold and collar gaskets for leakage.
|
|
| Clogged column. |
- Lysate was too viscous to pipette easily.
|
- Dilute lysate with Wizard SV Lysis Buffer until it becomes
easy to pipette. Then apply the entire lysate to the column.
|
- Too much tissue sample was used in the lysate preparation.
|
- The maximum recommended weight of mouse tail tissue used
for lysate preparation is 20 mg.
|
- Too many cells were being processed.
|
- The maximum recommended amount of tissue culture cells
being processed on the column
membrane is 5 X 106 cells.
|
- Tissue lysates were too viscous when cooled.
|
- Process lysates at 55°C incubation. If samples have
cooled, it may be difficult to purify due to viscosity.
Incubate at 55°C for 1 hour and continue with the purification.
|
- Differential vacuum pressure in filter plate.
|
- 1. Manually place Vacuum Assist film from Whatman, Cat.
# 7705-0112, on the filter plate during vacuum steps to
ensure even vacuum pull in all 96 columns.
|
|
| Uneven reagent distribution. |
- Uneven multi-channel pipetting head.
|
- Inspect the levelness of the
multi-channel pipetting head by lowering the pod via manual
control to touch a 1X1 ALP on the deck. Uneven contact of
the head to the ALP signifies a tilted head. Contact a service
engineer to relevel the pipetting head before operating
the Biomek® FX.
|
| |
- Need to refurbish head and
inspect tips for blocks.
|
- Loading partially filled tip boxes.
|
- Only full tip boxes can be used. Partially filled tip
boxes creates uneven insertion force on the
multi-channel pipetting head, which may cause damage and
pipette inaccuracy to the multichannel pipetting head.
|
|
Reagents are not
delivered correctly. |
- Incorrect positioning of reagents.
|
- Consult the instrument setup screen to ensure reagents
are placed in the appropriate location.
- Check to make sure that lysis buffer, wash solution, and
nuclease-free water are added to the appropriate labware
and placed in the correct deck position.
|
- Reagents were not delivered to the desired position in
wells.
|
|
|
| RNA contamination. |
- RNase A was not added to the tissue lysate digestion
solution.
|
- Add 2 uL of RNase A solution to final eluate and incubate
at room temperature for 10 minutes.
|
- RNA was co-purified with genomic DNA from tissue culture
cells.
|
- Add 2 uL of RNase A solution to final eluate and incubate
at room
temperature for 10 minutes.
|
|
| Contamination of wells. |
- Outside sources introduced during sample preparation.
|
- Make sure that Biomek® FX and reagents are free of
contaminants.
|
|
* All trademarks are the property of their respective owners. Where applicable,
the PCR process is covered by patents owned by Roche Molecular Systems,
Inc., and F. Hoffman-LaRoche, Ltd.
eLabNotebook
Sitemap
|