Specifications | Ordering Information

Specifications
|
Avanti
J-30I |
Avanti
J-26 XP |
Avanti
J-26 XPI |
Avanti
J-E |
| Maximum Speed (rpm) |
30,000 rpm |
26,000
rpm |
21,000 rpm |
Maximum Force
(RCF x g) |
110,500 x g |
82,000
x g |
53,000 x g |
| Maximum Capacity |
4 liters |
6 liters |
4 liters |
| Single Tube Volume Range |
1.5 - 1,000
mL |
| Rotor I.D. |
Manual |
| Continuous Flow |
Yes |
| Zonal |
Yes |
No |
Yes |
No |
| Programmable |
30, 2-Step Programs |
No |
30, 2-Step Programs |
9 single-step programs |
| Accel/Decel Rates |
12/13 |
2/3 |
12/13 |
2/3 |
| Delay Start |
Yes |
No |
Yes |
Yes |
| Run Set Times |
1 Min-99:59 Hr., Hold,
or w²t |
1 Min-3 Hr., Hold |
1 Min-99:59 Hr., Hold,
or w²t |
1 Min-99:59
Hr., Hold, or w²t |
Set Temp Range
(1° increments) |
-20°C to 40°C |
-10°C to 40°C |
-20°C to 40°C |
| Temp Control |
±2° of
set temperature |
| Speed Control |
±10 rpm of set speed
or 0.15%, whichever is greater |
±10 rpm
of set speed or 0.1%, whichever is greater |
| Ambient Temperature Range |
15°C to 40°C |
| Drive Type / Cooling |
SR†
Drive / Air Cooled |
| Refrigeration System |
Non-CFC,
non-ozone depleting refrigerant |
| Power Usage / Heat Output |
2.0 kW
/ 6900 Btu/Hr |
| Sound Level |
< 57
dbA (3 ft. from instrument at maximum speed) |
< 65 dbA |
Dimensions
H x D x W |
865 mm x 876 mm x 711
mm
(34.1 in. x 34.5 in. x 28 in.) |
914 mm x 800 mm x 635 mm (36.0 in.
x 31.5 in. x 25.0 in.) |
| Weight |
310 kg
(680 lb.) |
290 kg
(640 lb.) |
267.5 kg
(589 lb.) |
back to top

Ordering Information
* Beckman Coulter Elutriator Rotors are for research and development
use only. They have not been designed for use in diagnostic or
therapeutic procedures.
back to top
†SR Drive is manufactured by Beckman Coulter under license from
Switched Reluctance Drives, Ltd., Leeds, UK.