Beckman Coulter, Inc.
 

Products

   
   
 
 

nanoVette Microliter Measurement Cell Specifications

Printer Friendly


Related Product Information

Performance Specifications

  Sample
Specific
Factor
1.0 mm lid
(virtual dilution factor 10)
[ng/µL]
0.2 mm lid
(virtual dilution factor 50)
[ng/µL]
Total Detection
Range
[ng/µL]
dsDNA
50
10 - 1,000
200 - 5,000
10 - 5,000
ssDNA
37
7 - 740
150 - 3,700
7 - 3,700
ssRNA
40
8 - 800
160 - 4,000
8 - 4,000
Oligo
30
6 - 600
120 - 3,000
6 - 3,000

Note: The ranges for ssDNA, ssRNA, and Oligo are based on the results from dsDNA.

Physical Specifications

  A44097 A44098 A44100
  nanoVette with 0.2 mm pathlength lid nanoVette with 1.0 mm pathlength lid nanoVette with 0.2 and 1.0 mm pathlength lids
Width 12.5 mm
Depth 12.5 mm
Height 53 mm DU® 800
54.5 mm DU 730*
Pathlength/
Virtual Dilution
0.2 mm / 1:50 1.0 mm / 1:10 0.2 mm / 1:50
1.0 mm / 1:10
Sample Volume 0.7 µL - 4.0 µL (0.2 mm lid) 3.0 µL - 5.0 µL (1.0 mm lid) 0.7 µL - 4.0 µL (0.2 mm lid)
3.0 µL - 5.0 µL (1.0 mm lid)

* The DU 730 requires the addition of a 1.5 mm height adjustment step, included with the nanoVette

Questions regarding the nanoVette microliter cell? Please email nanoVette@beckmancoulter.com

Additional Information:

 
 
Home | Products | Customer Support | Resource Center
Employment | Our Company | Contact Us | Site Map