Beckman Coulter, Inc.
 

Products

   
   
 
 

DU® 800 UV/Vis Spectrophotometer Specifications

Printer Friendly


Related Product Information

The following are "guaranteed" and not "typical" specifications, as sometimes claimed by other companies. 

Performance Specifications

 Wavelength Range  190 to 1100 nm
 Wavelength Setability  0.1 nm Increments
 Wavelength Accuracy
     - At 656.1 nm
     - Full Range
 
 + 0.2 nm
 + 0.5 nm
 Wavelength Repeatability
     - At 656.1 nm
     - Full Range
 
 + 0.1 nm
 + 0.2 nm
 Spectral Bandwidth (from 200 to 680 nm)  < = 1.8 nm
 Photometric Readout  -0.300 to 3.000A or 0.0 to 200%T
 Photometric Accuracy
 (at 1A with NIST 930D solid filter at 546 nm)
 + 0.005A
 RMS Noise
 (at 0A, average of 10 standard deviations of 10
 readings at 0.05 sec intervals, at 500 nm)
 < 0.0002A RMS
 Stray Light
 (measured using NaI at 220nm, per ASTM
 E387-84)
 < 0.05%T
 Stability
 (0A, constant ambient conditions, measured for
 one hour at 340 nm after one hour warmup)
 < 0.003A/hr
 RMS Baseline Flatness
 (from 200 to 900 nm, at 0 A)
 + 0.001A RMS
 Scanning Speeds  120, 240, 600, 1200, 2400 nm/min

Physical and Environmental Specifications

 Width  69 cm (27 inches)
 Height  53 cm (21 inches)
 Depth  26 cm (10 inches)
 Weight  37 kg (81 pounds)
 Line Voltage  110 - 240V +10%
 Frequency  50/60 Hz
 Power  200 watts typical
 Temperature Operating Range  +15 to 40ºC (59 to 104ºF)
 Humidity  < 85% maximum relative humidity, not to
 exceed 32.5?C WBT

The physical specifications apply to the optical bench, which is the DU 800 Spectrophotometer without a computer.